QUICK GUIDE

MEASUREMENT

Ask the patient to relax and look 

1. 

straight 

ahead at a specific point

 to keep the eye 

steady during the measurement.

Bring the tonometer near the patient’s eye. 

2. 

Check that the 

central groove is in a horizon-

tal position

. It’s not necessary to look at the 

display during the measurement since the 

device gives sound signals about errors etc.

Keep the distance from the tip of the probe 

3. 

to the patient’s cornea at 

4 to 8 millimeters 

(1/6-1/3 inch)

. If necessary, adjust the 

forehead support.

Press the measuring button lightly. The tip of 

4. 

the 

probe should make contact with the 

central cornea

.

Six measurements are made. Once the six 

5. 

correct measurements are done, the IOP will 

be shown on display.

Please refer to the manual for additional in-

6. 

formation.

ONE SHORT BEEP

  

Indicates a correct single measurement. 

DOUBLE BEEP

  

––

Indicates an incorrect measurement.

(See error codes from manual).

LONG BEEP

Indicates that six succesfull measure-

ments are done and IOP result is available 

at the display.

Press either selector button until the dis-

play shows ‘End‘. Press the measuring but-

ton for two seconds - the display will show 

‘byE’ and the tonometer will switch off.

If the tonometer is not used for two

minutes, it will automatically switch off.

4–8 mm

(1/6–1/3”)