PARTS OF THE TONOMETER

CLEANING THE PROBE BASE

The probe base can be reused after careful cleaning. Make sure 
that the probe base is completely dry before using it.

1.

    

Unscrew the probe base collar and carefully remove the probe 
base from the tonometer.

2.   Invert the probe base over a container filled with 70-100 % 

alcohol solution, allowing the probe base to drop into the 
container and let soak for 5 - 30 minutes.

3.   Dry the probe base by blowing some compressed air into it.

4.   Carefully reinsert the completely dry probe base, screw the 

collar back into the tonometer.

Please refer to the manual for more detailed instructions of 
changing and cleaning the probe base.

CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID 
UNNECESSARY SERVICE.

www.icaretonometer.com