Green probe base light indication

ONE SHORT BEEP 

Indicates a correct single measurement. 

DOUBLE BEEP 

Indicates an incorrect single measurement. (See error messages from full manual).

LONG BEEP 

Indicates that six succesful measurements are done and IOP result is available 

at the display.

ENGLISH

Icare ic100 tonometer Quick Guide

Step 2.

  

Keep the probe horizontal and pointing perpendicularly to the center of the cornea. The distance 

from the tip of the probe to the patient´s cornea should be 4-8 mm (5/32 - 5/16 inch). The probe 

base light must be green.

Step 3.

 Press the Measure button to measure IOP.

Probe will contact the eye during measurements: Press and hold the button to collect 6 measurements 
in a row OR Press the button briefly 6 times to collect 1 measurement at the time.

DISPLAY DURING MEASUREMENTS

Before

After the second 

measurement

After the sixth 

measurement

After the sixth 

measurement but with 

too much variation

After the sixth 

measurement but with 

acceptable variation

REPEAT