Green probe base light indication
ONE SHORT BEEP
Indicates a correct single measurement.
DOUBLE BEEP
Indicates an incorrect single measurement. (See error messages from full manual).
LONG BEEP
Indicates that six succesful measurements are done and IOP result is available
at the display.
ENGLISH
Icare ic100 tonometer Quick Guide
Step 2.
Keep the probe horizontal and pointing perpendicularly to the center of the cornea. The distance
from the tip of the probe to the patient´s cornea should be 4-8 mm (5/32 - 5/16 inch). The probe
base light must be green.
Step 3.
Press the Measure button to measure IOP.
Probe will contact the eye during measurements: Press and hold the button to collect 6 measurements
in a row OR Press the button briefly 6 times to collect 1 measurement at the time.
DISPLAY DURING MEASUREMENTS
Before
After the second
measurement
After the sixth
measurement
After the sixth
measurement but with
too much variation
After the sixth
measurement but with
acceptable variation
REPEAT