www.icaretonometer.com
32
Icare HOME
Icare HOME tonometer Instruction manual for health care professionals
English
Guidance and manufacturer’s declaration IEC 60601-1-2: Edition 4.0; 2014-02 – Electromagnetic immunity
Icare HOME (TA022) is intended for use in a home healthcare environment with
electromagnetic characteristics specified below. The customer or the user of
the Icare HOME (TA022) should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment-Guidance
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
0,15 MHz – 80
MHz
6 V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 V
6 V
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the Icare HOME (TA022), including
cables specified by the manufacturer.
Otherwise, degradation of the performance of
this equipment could result.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz
10 V/m
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the Icare HOME (TA022) including cables
specified by the manufacturer. Otherwise,
degradation of the performance of this
equipment could result.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Icare HOME tonometer Instruction manual for health care professionals
English
www.icaretonometer.com
33
Icare HOME
Guidance and manufacturer’s declaration IEC 60601-1-2: Edition 4.0; 2014-02 – Electromagnetic immunity
Icare HOME (TA022) is intended for use in a home healthcare environment with electromagnetic
characteristics specified below. The customer or the user of the Icare HOME (TA022) should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment-Guidance
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz 27 V/
m; PM 50%; 18 Hz
430 - 470 MHz 28 V/m;
(FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz 9 V/m;
PM 50%; 217 Hz
800 - 960 MHz 28 V/m;
PM 50%; 18 Hz
1700 - 1990 MHz 28
V/m; PM 50%; 217 Hz
2400 - 2570 MHz 28
V/m; PM 50%; 217 Hz
5100 - 5800 MHz 9 V/
m; PM 50%; 217 Hz
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the
device1), including cables specified by the
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Icare HOME tonometer Instruction manual for health care professionals
ENGLISH
32
www.icare-usa.com