22
Door adjustment
Under normal circumstances, the chamber door does not
require adjustments. However, if the seal fails (resulting in
steam leaking from the front of the chamber), you may
adjust it.
Open the door, insert your finger into the bottom of the
door, and pull the ring to unlock the mechanism. Turn the
lid counterclockwise to tighten it. If it is too tight, you may
also turn the lid clockwise to loosen it.
Replacement of the door seal ring
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:
Press in the top and bottom of
the door seal.
Press in the left and right sides
of the door seal.
Press the remaining sections of
the seal.
Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.
The drain valves
To drain the water from the used water tank or to fill the tank from an external
container, insert the connector attached to the silicone tube until a click is
heard. To disconnect the connector, press the metal button.