17
Once the cycle has been programmed a
message icon will appear on the main screen.
Tap Message to review the parameters of the
programed cycle or/and to cancel it.
Test Programs
Helix Test
Tap Test from the main menu, select Helix test, in the
screen will appear the information of selected Test
program as the pressure, temperature, time (this may
vary depending on the environmental conditions and
the load), vacuum times, Sterilization time (holding
time) and dry time.
Following the instructions of the test manufacturer.
Put the Helix test into the chamber, then close the
door and Tap “ ”
After finishing the cycle, you may check the indicator
and evaluate the result according with the instructions
of the test manufacturer.
B&D test
Tap “Test” from the main menu, select “B&D test”, in
the screen will appear the information of selected Test
program as the pressure, temperature, time (this may
vary depending on the environmental conditions and
the load), vacuum times, Sterilization time (holding
time) and drying time. Following the instructions of the
test manufacturer.
Put the B&D test into the chamber, then close the
door and Tap “ ”
After finishing the cycle, you may check the indicator
and evaluate the result according with the instructions
of the test manufacturer.