INTRODUCTION

INTRODUCTION

      

Thank you for selecting IC Sciences Corporation’s Blood Pressure Monitor 

(LS810-B). The monitor features blood pressure measurement, pulse rate 

measurement and the data storage. The design provides you with two  

years of storage.

This  manual   contains

    important

     safety

      and

       care 

        informat  

 iion

       and

 

 

 

provides step-by-step instructions for using the product.

Please read this user manual carefully and thoroughly before use.

FEATURES:

Systolic Blood Pressure
Diastolic Blood Pressure

Pulse Rate
Memory: Up to 60 records

This product uses the Oscillometric Measuring Method to detect blood pressure. 

Before every measurement, the unit establishes a “zero point” equivalent to the 

atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit 

detects pressure oscillation generated by beat-to-beat pulsatile, which is used 

to determine the systolic pressure and diastolic pressure as well as pulse rate. 

The device also compares the longest and the shortest intervals of detected 

pulse wave with the average value, and then calculates the standard deviation.

The monitor will light up a warning symbol when the calculated standard 

deviation is larger than or equal to 15. 

Measurement Principle

General Description

The below signs might be in the user manual, labeling or other components. 

They are the requirement of standard and using. 

Safety Information

Symbol for “THE OPERATION 

GUIDE MUST BE READ”

Symbol for “MANUFACTURER”

Symbol for “SERIAL NUMBER”

Symbol for “TYPE BF APPLIED 

PARTS”

Symbol for “DIRECT CURRENT”

Symbol for “ENVIRONMENT 

PROTECTION - Waste electrical products 

should not be disposed of with 

household waste. Please recycle where 

facilities exist. Check with your local 

authority or retailer for recycling advice”

The Bluetooth Combination Mark

Symbol for “MANUFACTURE 

DATE”

2

3

Indications for use

The IC Sciences Corporation Blood Pressure Monitor is a digital 

monitor intended for use in measuring blood pressure and heartbeat rate with 

wrist circumference ranging from 13.5 cm to 21.5 cm ( about 5-8.5 inches ). 

It is intended for adult  indoor use only.

Symbol for “ Class II Equipment” 

CAUTION

* It is intended for adult indoor use only. Pregnant women, pre-eclamptic patients,

and severely obese patients should not use the device. If you need to measure your blood  

pressure, please consult your physician.

* This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is

not intended for use on extremities other than the wrist or for functions other than obtaining a

blood pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood

pressure. Please start or end medical treatment basing solely on physician’s treatment advice.

* If you  are taking medication, consult your physician to determine the most appropriate time for

your measurement. Never change a prescribed medication without your physician’s consent.

* This unit is not suitable for continuous monitoring during medical emergencies or operations.

Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen, and even purple due

to a lack of blood.

* If the pressure of the cuff exceeds 40 kPa (300 mmHg), the unit will automatically deflate. If the

cuff doesn’t deflate when its pressure exceeds 40 kPa (300 mmHg), detach the cuff from the wrist 

and press the START/STOP button to stop inflation.

* Do not use the monitor under the conditions of strong electromagnetic field (e.g. medical RF

equipment) that radiates interference signal or electrial fast transient/ burst signal.

* The maximum temperature that the applied part can be achieved is 42.5 

℃  

(108.5 F) while the

environmental temperature is 40

℃ 

(104 F)

.

* The device is not AP/APG equipment. It is not suitable for use in the presence of a flammable

anesthetic mixture with air (or oxygen, nitrous oxide).

* Please keep the unit out of reach of infants, children or pets, since inhalation or swallowing of

small parts is dangerous or even fatal.

* Please use ACCESSORIES and detachable parts specified / authorised by MANUFACTURER.

Otherwise, it may cause damage to the unit or danger to the user / patient.

* The patient is an intended operator. The patient can measure, transmit data and charge battery

under normal circumstances and maintain the device and its accessories according to the user

manual.

* The blood pressure monitor, and the cuff are suitable for use within the patient

environment. If you are allergic to dacron or plastic, please don’t use this device.

* The device is not intended for PATIENT transport outside a healthcare facility.

*This device cannot be used with HF surgical equipment at the same time.

* There is a PTC current limiter in the monitor, which specification is 8V and 0.5A. When the voltage

and current exceed the limiting value, the monitor will stop working.

   

   

 

 

 

 

* If Luer lock connectors are used in the construction of tubing, there is a possibility that they might 

be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood 

vessel.

* Be careful of strangulation due to cables and hoses, particularly due to excessive length.

Symbol for “Including RF

transmitter”

     Readings taken by the LS810-B are equivalent to those obtained by a trained

observer using the cuff and stethoscope auscultation method.

Summary of Contents for LS-810-B

Page 1: ...al for further reference in case you have problems Thank you for selecting the IC Sciences Corporation Blood Pressure Monitor LS810 B Wrist Type Version 1 0 MEM SET FCC ID OU9LS810 B01 Distributed by Informed Clinical Sciences Corporation 1 Joy Street Boston USA MA 02108 Download and install My BP Coach from iTunes Store or Google play My BP Coach ...

Page 2: ...get from the hospital different from home Is the result the same if measuring on the right wrist TROUBLESHOOTING 20 SPECIFICATIONS 21 THE MATCHED COMPONENTS 22 CONTACT INFORMATION 22 Safety Information Power Supply and Charge Power Activate Your Blood Pressure Monitor Setting Date and Time Positioning the Cuff Start Measurement Data Management 14 Recall the Records Delete the Records COMPLIED STAN...

Page 3: ... pressure measurement Do not confuse self monitoring with self diagnosis This unit allows you to monitor your blood pressure Please start or end medical treatment basing solely on physician s treatment advice If you are taking medication consult your physician to determine the most appropriate time for your measurement Never change a prescribed medication without your physician s consent This unit...

Page 4: ... the power Error Memory Recalling the history records Bluetooth Successful Bluetooth Connection IHB Detector Irregular Heartbeat Detector Error 5 SYSTOLIC DIASTOLIC MEM UP BUTTON SET DOWN BUTTON START STOP BUTTON CUFF Type BF applied part PULSE RATE TIME USB Interface Monitor Components List 1 Blood Pressure Monitor LS810 B 2 USB Cable 3 User Manual MEM SET Component list of pressure measuring sys...

Page 5: ...can cause the battery to explode 4 You can only use the IC Science s authorized USB cable to charge the power You cannot use the blood pressure monitor during charging 5 During charging the blood pressure monitor displays the symbol When the charging is finished please unplug the device 6 When charging do not touch the charging connector and the patient simultaneously 7 Setting Date and Time 2 As ...

Page 6: ...ch from the bottom of the wrist 5 Fasten the wrist cuff around your wrist leaving no extra room betweenthe cuff and your skin If the cuff is too loose the measurement will not be accurate Rest for 5 minutes before measuring your blood pressure Wait at least 3 minutes between measurements This allows your blood circulation to recover For a meaningful comparison try to measure under similar conditio...

Page 7: ...itor and it will complete the measurement process Start Measurement CAUTION Interference may occur in the vicinity of equipment marked with the following symbol And LS810 B may interfere in the vicinity of electrical equipment Keep the monitor at least 20 centimeters 8 inches away from the human body especially the head when the data transmission is proceeding after measurement To enable the data ...

Page 8: ...AF premature ventricular beats and peripheralarterial disease PAD should avoid using the unit whenever possible 1 When using this device please pay attention to the following situations which may interrupt blood flow and influence blood circulation of the patient thus cause harmful injury to the patient too frequent and consecutive multiple measurements the application of the CUFF and its pressuri...

Page 9: ...ord 60 is dropped from the list CAUTION DATA MANAGEMENT 14 When you did not obtain an accurate measurement you can clear all the results by following the steps below Delete the Records 1 Under Memory Recalling Mode press and hold both the MEM button and the SET button for 3 seconds 2 The LCD will display dEL dONE indicating that memory clearing is complete Then it will shutdown To cancel the clear...

Page 10: ...it is in proper working conditions before use Don t service or maintain while the device is in use 2 If you have any problems with this device such as setting up maintaining or using please contact Informed Clinical Sciences Do not open or repair the device yourself 3 Please report to Informed Clinical Sciences if any unexpected operation or events occur 4 Cleaning A dusty environment may affect t...

Page 11: ...as reached a dangerous point CAUTION SYS 135 mm Hg DIA 85 mm Hg Blood Pressure Category Normal Prehypertension High Blood Pressure Hypertension Stage 1 High Blood Pressure Hypertension Stage 2 Hypertensive Crisis Emergency care needed Systolic mmHg upper Diastolic mmHg lower less than 120 120 139 140 159 160 or higher Higher than 180 and or or or or less than 80 80 89 90 99 100 or higher Higher th...

Page 12: ...splay moder Digital LCD V A 46 5x36 5mm Measurement mode Oscillographic testing mode Measurement range Accuracy Normal working condition Storage transportation condition Measurement perimeter of the wrist About 13 5cm 21 5cm Net Weight Approx 110g External dimensions Attachment Approx 79 8 72 5 13 2mm USB cable user manual Mode of operation Continuous operation Degree of protection Type BF applied...

Page 13: ... Symbols to be used with medical device labels labelling and information to be supplied General requirements EN 1041 2008 Medical equipment manufacturers to provide information IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC EN 60601 1 2 2007 Medical electrical equipment Part 1 2 General requirements for basic safety ...

Page 14: ... This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment This transmitter must not be co located or operating in conjunction with any other antenna or transmitter EMC Guidance 24 25 1 The Blood Pressure Monitor needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPAN...

Reviews: