![IBT morph lock Installation And Setup Manual Download Page 17](http://html.mh-extra.com/html/ibt/morph-lock/morph-lock_installation-and-setup-manual_609365017.webp)
17
Before fitting the user with the prosthesis, test the prosthesis first.
Once all the components are connected and inside the prosthesis, connect
the hand and turn the FlexCell battery system on. Wait 5 seconds.
morph
lock
will vibrate within 5 seconds to signal that it has received power and is
ready to use.
Tap the electrodes with your finger tip and confirm that the hand opens and
closes appropriately. If the signals are reversed (i.e. close electrode opens the
hand), connect to the hand’s software and switch the EMG signals.
Approach a tag to the wrist area of the prosthesis where the
morph lock
antenna resides.
morph lock
will vibrate twice to indicate that it has detected
the tag and locked the hand. Tap the electrodes to confirm that the hand
does not respond to EMG. Approach another tag and
morph lock
will vibrate
once to indicate that it has unlocked the hand. Tap the electrodes to confirm
that the hand now responds to EMG.
Familiarize yourself with
morph lock
’s detection range by approaching tags
to other areas around the wrist.