40
41
8. In case problems occur
9. Technical Data
Manufacturer:
Well-Life Healthcare Limited
1Fl., No. 16 Lane 454, Jungjeng Rd.
Yunghe City, Taipei City
Taiwan, R.O.C
Type of device:
Device for transcutaneous electrical nerve
stimulation (TENS)
Classification according to 93/42/EWG:
Class IIa
Model:
WL-2402
Maximum permissible ambient temperature:
10°C - 45°C
Maximum permissible relative humidity:
30% - 75%
Maximum permissible air pressure:
700 hPa - 1060 hPa
Battery:
Coin cell Type CR2032, 3 V,
non-rechargeable
Max. output of electrodes:
50 mA
Pulse width:
150 - 250 µs
Pulse rate:
30 - 60 Hz
Operating voltage:
3 V
Weight:
10 g (w/o Battery)
Dimensions (mm):
W:55 x D:15 x H:70
Problem
Possible Cause
Solution
The Slim
Tens is switched
on, but does not function.
The battery is not inserted.
Insert the battery
.
The battery is empty
.
Replace the battery
.
The Slim
Tens is defective.
Mail your Slim
Tens to our service
department, see
Service and W
ar
-
ranty
.
or
Dispose of the Slim
Tens, see
Dis
-
posal and Environmental Protection
The stimulation is too weak.
The battery is low
.
Replace the battery
.
The patch does not adhere
any longer
.
The patch is dried out.
Replace the patch.