6-16
Chapter 6, COMPUTER OPERATION
AXO
Auto-Offset
The X and Y output offsets are turned on and set to levels giving zero X and Y
outputs. Any changes in the input signal then appear as changes about zero in the
outputs.
EX [n]
Output expansion control
Expands X and/or Y outputs by a factor of 10. Changes meter, CH1 and CH2
outputs full-scale to ±10 % if X or Y selected. The value of n has the following
significance:
n
Expand mode
0
Off
1
Expand X
2
Expand Y
3
Expand X and Y
CH n
1
[n
2
]
Analog output control
Defines what outputs appear on the rear panel CH1 and CH2 connectors according
to the following table:
n
2
Signal
0
X %FS
1
Y %FS
2
Magnitude %FS
3
Phase 1:- +9 V = +180°, -9 V = -180°
4
Phase 2:- +9 V = 360°, - 9 V = 0°
5
Noise %FS
6
Ratio:- (1000 × X)/ADC 1
n
1
is compulsory and is either 1 for CH1 or 2 for CH2
6.4.05 Instrument Outputs
X[.]
X output
In fixed point mode causes the lock-in amplifier to respond with the X demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the X demodulator
output in volts or amps.
Y[.]
Y output
In fixed point mode causes the lock-in amplifier to respond with the Y demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the Y demodulator
output in volts or amps.
Summary of Contents for 7220
Page 8: ...v i TABLE OF CONTENTS...
Page 12: ...1 4 Chapter 1 INTRODUCTION...
Page 76: ...5 36 Chapter 5 FRONT PANEL OPERATION...
Page 124: ...D 2 Appendix D CABLE DIAGRAMS Figure D 2 Interconnecting RS232 Cable Wiring Diagram...
Page 146: ...E 22 Appendix E ALPHABETICAL LISTING OF COMMANDS...
Page 154: ...Index 6 INDEX...