E-15
Appendix E, ALPHABETICAL LISTING OF COMMANDS
SEN[.] [n]
Full-scale sensitivity control
The value of n sets the full-scale sensitivity according to the following table,
depending on the setting of the IMODE control:
n
full-scale sensitivity
IMODE=0
IMODE=1
IMODE=2
4
20 nV
20 fA
n/a
5
50 nV
50 fA
n/a
6
100 nV
100 fA
n/a
7
200 nV
200 fA
n/a
8
500 nV
500 fA
n/a
9
1 µV
1 pA
n/a
10
2 µV
2 pA
20 fA
11
5 µV
5 pA
50 fA
12
10 µV
10 pA
100 fA
13
20 µV
20 pA
200 fA
14
50 µV
50 pA
500 fA
15
100 µV
100 pA
1 pA
16
200 µV
200 pA
2 pA
17
500 µV
500 pA
5 pA
18
1 mV
1 nA
10 pA
19
2 mV
2 nA
20 pA
20
5 mV
5 nA
50 pA
21
10 mV
10 nA
100 pA
22
20 mV
20 nA
200 pA
23
50 mV
50 nA
500 pA
24
100 mV
100 nA
1 nA
25
200 mV
200 nA
2 nA
26
500 mV
500 nA
5 nA
27
1 V
1 µA
10 nA
Floating point mode can only be used for reading the sensitivity, which is reported in
volts or amps. For example, if IMODE = 0 and the sensitivity is 1 mV the command
SEN would report 18 and the command SEN. would 1.0E-03. If IMODE was
changed to 1, SEN would still report 18 but SEN. would 1.0E-09.
SLOPE [n]
Output low-pass filter slope (roll-off) control
The value of n sets the slope of the output filters according to the following table:
n
slope
0
6 dB/octave
1
12 dB/octave
2
18 dB/octave
3
24 dB/octave
SRATE[.] [n]
Oscillator frequency sweep step rate
Sets the sweep rate in time per step in the range 50 ms to 1000 s, in 5 ms increments.
Summary of Contents for 7220
Page 8: ...v i TABLE OF CONTENTS...
Page 12: ...1 4 Chapter 1 INTRODUCTION...
Page 76: ...5 36 Chapter 5 FRONT PANEL OPERATION...
Page 124: ...D 2 Appendix D CABLE DIAGRAMS Figure D 2 Interconnecting RS232 Cable Wiring Diagram...
Page 146: ...E 22 Appendix E ALPHABETICAL LISTING OF COMMANDS...
Page 154: ...Index 6 INDEX...