1.1.2
The software of the equipment provides:
−
the special measuring algorithm. According to this algorithm the unit estimates and
corrects the measured data every 10 seconds repeatedly. In constant ionization
radiation field the readings taken by the unit are being continuously corrected and
averaged to reduce the statistical error.
Software allows to display:
−
the
ambient dose and the ambient dose rate;
−
the radiation environment time based historical graphs;
−
the current time and environment temperature;
−
the battery charge level;
−
the measurement cycle dynamic graph;
−
the dose accumulated during the current measurement or the summarized value of
the dose accumulated for all the measuring times).
The software also gives the opportunity of:
−
non-volatile storing of the accumulated dose;
−
selecting the unit measures (roentgen, Sievert);
−
setting and changing the alarm sound of the ambient dose rate threshold;
−
audiovisual warning of the ambient dose rate threshold excess;
−
setting the data display language (Russian, English);
−
performing the audiovisual signals at defined frequency that helps to find out the
sources of the radiation quickly;
−
setting the screen options (color skin of the menu, brightness and the screen
lighting off time to spin out the energy);
−
setting the sound options (switching on and off the whole sound, the sound of
touching buttons, alarm sound of the ambient dose rate threshold, the particle
detection sound).
1.1.3. The device is intended to use at the temperature ranging from -5
о
С to +40
о
С, the
relative moisture below 98% at the temperature of +25
о
С and at the atmosphere pressure of 84-
106,7 kPa.
1.2. Technical features
1.2.1. The Figure below illustrates the physical configuration of the product (Fig.1).
Fig.1 – Physical configuration