Replacing the battery
If, after normal recharging, the battery cannot sustain a full treatment session, recharge again, as it is
probably approaching the end of its useful life. If even after a second recharge the battery still cannot
complete a treatment session, follow the steps here below:
•
Contact the manufacture or authorized dealer for original spare part;
•
Open the battery compartment on the back side of MAG1000, disconnect the red/black
wire from connector and remove the exhausted battery;
•
Insert new battery and connect the red/black wire to the plug;
•
Close battery compartment
WARNING: for disposal exhausted battery follow the instructions on chapter “Disposal”. Do not
open or burn battery. Do not short circuit poles. Keep battery away from sparks or naked flames. In
the event of internal electrolyte coming into contact with skin or garments, wash immediately with
water. In the event of electrolyte coming into contact with eyes, rinse thoroughly and seek medical
attention.
Maintenance
Cleaning
Clean the equipment from the dust using a dry soft cloth.
Resistant stains can be removed using a sponge soaked in solution of water and alcohol (20%).
Carriage and storage
Carriage precautions
MAG1000 is a portable device, so it does not need any particular carriage precautions.
However we recommend to put away MAG1000 and its accessories in their own bag after every
treatment.