MNPG127-02 

Pag. 6 

 

! WARNINGS! 

-

 

The perfect functionality of the device is guaranteed in accordance with the 
rules  of  installation  and  of  use  included,  only  with  original  accessories  and 
spare parts.  

-

 

If there are problems or installation difficulties, please contact the I.A.C.E.R. 
srl technical assistance department.  

-

 

The  correct  position  while  moving  the  machine:  the  apparatus  has  to  be 
moved  exclusively  by  gripping  it  with  both  hands  on  the  curved  profiles  of 
the lid. 

-

 

If you want to install an extern interlock circuit, contact exclusively qualified 
technicians and supply them the scheme correspondent to the room used for 
the  emission  of  the  treatment.  A  bad  installation  of  the  device  can  to 
generate serious ocular lesions.  

-

 

Before  connecting  the  cable  to  the  mains  plug,  check  that  the  equipment 
wasn’t  damaged  during  transport.  Ensure  that  the  power  supply 
specifications  on  the  mains  socket  correspond  with  the  information  on  the 
label attached to the back of the unit.  

-

 

The  electric  current  that  powers  the  unit  is  VERY  DANGEROUS.  Before 
connecting or disconnecting the power cable from the connecter on the unit, 
make sure it is plugged out from the mains socket.  

-

 

The power cable has an earthed plug for safety reasons.  

-

 

Only use with a mains socket suitable for use with earthed systems.  

-

 

The  equipment  should  only  be  connected  to  electrical  systems  that  fully 
comply with  
regulations.  

-

 

Check  the  integrity  and  the  existence  of  ground  conductor  if  extensions 
cables are used.  

-

 

Connect the equipment directly to the wall socket without using extensions. 
Failure  to  comply  with  these  warnings  may  result  in  dangerous  electrical 
discharges that could cause injury operators and compromise the functioning 
of the unit.  
 

USE 

-

 

The  laser  therapy  treatments  must  be  provided,  under  the  strict  control  of 
the  operator,  patients  conscious,  able  to  interact  with  the  operator  in 
response  to  stresses  transmitted  by  the  device;  in  case  of  default  to  the 
indications  given,  I.A.C.E.R.  srl  shall  not  be  consider  responsible  for  any 
accidents . 

-

 

The  use  of  the  controls  or  regulations  or  the  execution  of  different 
procedures from those specified in this user manual can cause the exposition 
to dangerous radiation. 

-

 

The  operator  has  the  responsibility  to  verify

 

that  the  issuing  head  remains 

well in contact  to the zone  of treatment,  to avoid the  emission to different 
zones from those to be treated.  

-

 

The laser radiation that outgoings from the device is dangerous: always use 
the appropriate glasses. Always avoid the exposition of the eyes to the direct 
or reflected laser beam. 

-

 

It  is  recommended  not  to  start  treatment  if  the  machine  is  not  in  perfect 
mechanical  condition  or  if  the  laser  does  not  present  characteristics  of 
approved for this purpose (see the specifications table).  

-

 

During  the  delivery  the  handpiece  must  be  positioned  in  contact  with  the 
part to be treated. After activating the handpiece through the pedal contact, 
avoid that it moves or is directed to different areas.  

-

 

THE PROBE MUST NEVER BE DIRECTED TO AREAS OF BODY SENSITIVE TO THE 
LASER RADIATION, FOR EXAMPLE THE EYES.  

-

 

ALWAYS AVOID THE EXPOSITION OF THE EYE TO THE DIRECT OR REFLECTED 
LASER BEAM.  

-

 

Do not leave the device switched on unattended, always switch off after use. 

-

 

Use  only  probes  accurately cleaned  and  disinfected  after each  treatment  in 
order to avoid environment and users contamination.  

 

The operator must pay attention to the necessity of a periodic maintenance 
(every  2  years)  of  the  probes/applicators.  IACER  srl  authorised  personnel 
should carry out such operations.  

 

It is absolutely forbidden in presence of anaesthetic inflammable substances 
and  in  full  oxygen  environment.  I.A.C.E.R.  will  be  not  responsible  if  these 
indications will be not complied.  

-

 

It  is  absolutely  forbidden  to  cover  the  ventilation slots:  such  an  action  may 
not allow the machine to work in safe conditions. In case of non-compliance 
with this indication, I.A.C.E.R.Srl will not be responsible for any accidents.  

-

 

It’s important to pay the attention of the operator to the necessity to verify 
the  correctness  of  the  electric  installation  of  device  before  activating  the 
supply switch. 

-

 

It is advisable to suspend the therapeutic treatment if it were to appear some 
disturbances during its emission.  

-

 

It’s strongly advised not to hold the device on in state of start without using 
the probe, it could overheat.  

 

Summary of Contents for LA500

Page 1: ...I A C E R Srl MNPG127 02 06 06 17 LASER THERAPY...

Page 2: ......

Page 3: ...E 8 INDICATIONS 8 CONTRA INDICATIONS 8 DEVICE DESCRIPTION 9 PATIENTS PREPARATION 9 HOW TO USE IT 9 Therapeutic indications 11 MAINTENANCE 13 HANDLES CLEANING 13 WORKING PROBLEM 13 ELECTROMAGNETIC INTE...

Page 4: ...important information and useful tips for operating the equipment CAUTIONS The CAUTION message appears before operations which if not correctly performed may cause damage to the machine and or its acc...

Page 5: ...t use the equipment in places where it might get wet Before operating the machine carefully check the correctness of the connections according to the instructions Do not use accessories other than the...

Page 6: ...ecution of different procedures from those specified in this user manual can cause the exposition to dangerous radiation The operator has the responsibility to verify that the issuing head remains wel...

Page 7: ...er production of ATP The applications of the I TECH LA500 laser produce a number of effects on the treated tissues 1 increase in hematic flow vasodilatation of capillaries and arteries 2 biostimulatio...

Page 8: ...ensitive to laser radiation and can be permanently damaged from direct or reflected laser beams The special safety glasses must be worn by both patient by the operator pregnancy the laser is contraind...

Page 9: ...of the patient for the therapy carry out the following operations 1 wash the skin using soap and mild soap or alcohol where the head of the laser will be positioned 2 dry well the skin HOW TO USE IT...

Page 10: ...ent time last 3 minutes Select the operation mode CONTINUOS or CYCLE using the buttons on the right side of the front panel picture below With CYCLE mode it is possible to select the ON time from 1 to...

Page 11: ...ition upper picture Take away the key from the machine to avoid the use by unauthorized personnel Please do not switch off the device unplugging the cable from the power socket to avoid any type of da...

Page 12: ...ITIS 0 3 66 sec 264 sec BAKER S CYST 0 3 66 sec 264 sec DE QUERVAIN SYNDROME 0 2 100 sec 400 sec OEDEMA 0 3 66 sec 264 sec HEMATOMA 0 3 66 sec 264 sec EPICONDYLITIS 0 4 50 sec 200 sec HERNIATED DISC 0...

Page 13: ...ent HANDLES CLEANING The laser handle is very weak and it needs a daily cleaning The following advices are absolutely important to not damage the lens and the laser probe It is suggested to 1 Remove t...

Page 14: ...that the socket works properly The power cable is not inserted correctly into the connector of the device Insert correctly the plug and the cable in the connector of the device The power cable is wor...

Page 15: ...B Output channels 1 Class of isolation parts applied according to the rule EN 60601 1 I BF Protection level against the liquids penetrations according to the rule EN 60601 1 IPX0 Command of execution...

Page 16: ...that the product shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognized centre of recycling of electrical and electronic equipment Appendix B...

Page 17: ...ctrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Page 18: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Page 19: ......

Page 20: ......

Page 21: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...

Reviews: