38 www.i-sens.com
www.i-sens.com 39
Note:
If the error messages persist, contact your authorised
i-SENS sales representative.
Message
The blood sample
has abnormally high
viscosity or
insufficient volume.
There is a problem
with the meter.
An electronic error
occurred during the
test.
Repeat the test after
inserting a new test
strip.
Do not use the meter.
Contact your
authorised i-SENS
sales representative.
Repeat the test with a
new test strip. If the
error message
persists, contact your
authorised i-SENS
sales representative.
What It Means
What To Do
This error message
may appear when
the wrong blood
β
-Ketone test strip is
used instead of the
KetoSens Blood
β
-Ketone test strip.
Repeat the test after
inserting a KetoSens
test strip.
Problem
Troubleshooting
• Check whether the test strip is
inserted with the contact bars facing
upwards. Check if the strip has been
inserted completely into the test
strip port.
• Check if the appropriate test strip
was used.
• Check whether the batteries are
inserted with the + side facing
upwards.
• Replace the batteries.
• Check if the confirmation window is
filled completely.
• Repeat the test with a new test strip.
• Repeat the test with a new test strip.
• Check the expiration date of the test
strip.
• Perform control solution test.
The display is blank
even after inserting
a test strip.
The test does not
start even after
applying the blood
sample on the strip.
The test result doesn't
match the way you
feel.
General Troubleshooting
Note:
If the problem is not resolved, please contact your
authorised i-SENS sales representative.
Summary of Contents for KetoSens