22 www.i-sens.com
www.i-sens.com 23
Note:
The Control Solution can be purchased separately. Please
contact your authorised i-SENS sales representative.
Caution:
The range printed on the test strip box is for the KetoSens
Control Solution only. It has nothing to do with your blood
β
-Ketone level.
Step 3
A test result will appear after the meter counts
down from 8 to 1.
When flagged, the result is stored in the meter’s
memory but it is not included in the averages.
Step 4
Compare the result displayed on the meter
to the range printed on the test strip box.
The result should fall within that range.
Used strips should be discarded safely in
appropriate containers.
Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip box. Repeat the
control solution test if the test result falls outside of this range.
Out of range results may occur in the following situations:
If results continue to fall outside the range printed on the test
strip box, the KetoSens Test Strip and Meter may not be working
properly. Do not use your system and contact your authorised
i-SENS sales representative.
Situations
• When the control solution bottle was
not shaken well,
• When the meter, test strip, or the control
solution were exposed to high or low
temperatures,
• When the first drop of the control
solution was not discarded or the tip of
the bottle was not wiped clean,
• When the meter is not functioning
properly.
• When the control solution is past the
expiration date printed on the bottle,
• When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months),
• When the control solution is
contaminated.
Do This
Repeat the
control solution
test by referring
to the 'Notes'
on page 20.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
Note:
The meter may switch off if the control solution is not
applied within 2 minutes of the
symbol appearing on the
screen. If the meter turns off, remove the strip, reinsert, and
start from step 1.
Control Solution Range
Control L :
0.4–1.0 mmol/L
Control M :
1.9–2.9 mmol/L
Summary of Contents for KetoSens