background image

34

35

Target Blood Glucose Ranges

Expected Values : 

The range of a normal fasting* blood glucose 

level for non-diabetic adults is between 70-99 mg/dL (3.9-5.5 
mmol/L). Two (2) hours after a meal, the range of a normal 
blood glucose level for non-diabetic adults is less than  
140 mg/dL (7.8 mmol/L).
*Fasting is defined as no caloric intake for at least eight (8) hours. 

Reference

American Diabetes Association. “Standards of Medical Care in 
Diabetes – 2016.” 

Diabetes Care

. January 2016; 39(1):S15, S100. 

Reminders

Time of day

Your target ranges

from your healthcare expert

Before breakfast
Before lunch or dinner
1 hour after meals
2 hours after meals
Between 2 a.m. and 4 a.m.

Transferring Test Results

Test results stored on 

CareSens II 

meter can 

be transferred from the meter to a computer 
using SmartLog software and cable. The meter 
screen displays 'Pc' when it is connected to 
the computer using the data cable.
For more information, contact your authorised 
i-SENS sales representative or visit  
www.i-sens.com.

Summary of Contents for CareSens II

Page 1: ...e monitoring through advanced biosensor technology Blood Glucose Monitoring System Owner s Booklet 2017 i SENS Inc All Rights Reserved PGA1E3350 REV3 02 2017 i SENS Inc 43 Banpo daero 28 gil Seocho gu...

Page 2: ...just 5 seconds with a small 0 5 L blood sample No part of this document may be reproduced in any form or by any means without the prior written consent of i SENS The information in this manual is cor...

Page 3: ...ter to Match Test Strips Checking the System Control Solution Testing Comparing the Control Solution Test Results Using the Lancing Device Preparing the Lancing Device Applying Blood Sample Discarding...

Page 4: ...code Temperature limitation Do not discard this product with other household type waste Serial number Glucose in blood samples reacts with the chemical in the test strip to produce a small electrical...

Page 5: ...e data transmission cable can be ordered separately Please contact your authorised i SENS sales representative CareSens II Blood Glucose Meter 1 Meter Display Test result symbol and simple message dis...

Page 6: ...play 1 2 3 4 5 6 7 8 12 11 10 9 Note It is recommended to check if the display screen on the meter matches the illustration on the page 8 every time the meter turns on Do not use the meter if the disp...

Page 7: ...tton to confirm your selection and progress to the next step 6 Setting the Year Press and release C button to adjust until the correct year appears Press the M button to confirm your selection and pro...

Page 8: ...results stored in the meter can be deleted Please note that if you select YES all the stored test results will be deleted and can not be restored After the unit is set press the M button to enter the...

Page 9: ...with the meter turned off Use the CareSens Check Strip when the CareSens II meter is used for the first time the battery is replaced thetestresultseemsinaccurateordoesnotreflecthowyoufeel the CareSens...

Page 10: ...areSens test strip package insert Keep the meter and testing supplies away from young children The test strip vial contains drying agents that are harmful if inhaled or swallowed and may cause skin or...

Page 11: ...rinted on their labels Compare the result displayed on the meter to the CareSens Control Solution range printed on the test strip vial Before using a new meter or a new vial of test strips you may con...

Page 12: ...he symbol will be displayed on the screen Confirm matching the code number on display and test strip vial If the code number does not match see Coding Meter to Match Test Strips on pages 17 18 2 Shake...

Page 13: ...the control solution test if the test result falls outside of this range Out of range results may occur in following situations If results continue to fall outside the range printed on the test strip...

Page 14: ...ays use a new sterile lancet Keep the lancing device clean Note Repeated puncturing at the same sample site may cause pain or skin calluses thick hard skin Choose a different site each time you test E...

Page 15: ...d sample using the lancing device Place the device against the pad of the finger as shown in the diagram The best puncture sites are on the middle or ring fingers Press the release button Remove the d...

Page 16: ...utomatically stored in the meter s memory If the test strip is removed after the test result is displayed the meter will automatically switch off after 3 seconds Discard used test strips safely in app...

Page 17: ...d hold the lancing device against site Wait until the skin surface under the lancing device changes color Then press the release button while continuing to apply pressure Keep holding the lancing devi...

Page 18: ...poglycemic shock or symptoms Note If the sample drop of blood runs or spreads due to contact with hair or with a line in your palm do not use that sample Try puncturing again in a smoother area Note P...

Page 19: ...tion Standards of Medical Care in Diabetes 2016 Diabetes Care January 2016 39 1 S15 S100 Reminders Time of day Yourtargetranges fromyourhealthcareexpert Before breakfast Before lunch or dinner 1 hour...

Page 20: ...from sources of strong electromagnetic radiation especially when measuring your blood glucose The CareSens II meter should be used only with CareSens strips Keep the meter in a cool and well ventilate...

Page 21: ...Humidity 10 90 Hematocrit 20 60 Specifications Message What It Means What To Do A used test strip was inserted Repeat the test with a new test strip The blood or control solution sample was applied b...

Page 22: ...strip The test does not start even after applying the blood sample on the strip The test result does not match the way you feel Check whether the test strip is inserted with the contact bars facing up...

Page 23: ...nts at clinic centers Slope 0 961 Y intercept 3 5 mg dL 0 2 mmol L Correlation coefficient r 0 995 Number of samples 115 Range tested 40 521 mg dL 2 2 28 9 mmol L Accuracy results for glucose concentr...

Page 24: ...rvice of the meter Any claim must be made within the warranty period The i SENS company will at its discretion repair or replace a defective meter or meter part that is covered by this warranty As a m...

Page 25: ...46 MEMO...

Reviews: