18
19
[3]
Press the
M
or the
C
button to select the
correct code number. Each time you press
the
M
button, the number will increase by
one, and each time you press the
C
button,
the number will decrease by one. Press and
hold the
C
button to scroll code numbers
down.
[4]
After selecting correct code number,
wait for three (3) seconds until the
symbol appears on display.
[5]
When the symbol appears,
meter is ready for testing.
You may check your meter and test strips using the
CareSens
Control Solution (control A and/or B).
The
CareSens
Control Solution contains a known
amount of glucose and is used to check that the
meter and the test strips are working properly.
The test strip vials have
CareSens
Control Solution
ranges printed on their labels. Compare the result
displayed on the meter to the
CareSens
Control
Solution range printed on the test strip vial.
Before using a new meter or a new vial of test strips,
you may conduct a control solution test following the procedure
on pages 21-22.
Notes:
•
Use only the
CareSens
Control Solution (available for
purchase separately).
•
Check the expiration date printed on the bottle. When you
first open a control solution bottle, record the discard date
(date opened plus three (3) months) in the space provided
on the label.
•
Make sure your meter, test strips, and control solution are at
room temperature before testing. Control solution tests must
be done at room temperature (20-25˚C/68-77˚F).
•
Before using the control solution, shake the bottle, discard
the first few drops and wipe the tip clean.
•
Close the control solution bottle tightly and store at a
temperature between 8-30˚C (46-86°F).