Understanding Error Messages
Message
What it Means
What To Do
A used test strip
was inserted.
Repeat the test with a new
test strip.
The blood or
control solution
sample was
applied before the
symbol
appeared.
Repeat the test with a new
test strip and wait until the
symbol appears before
applying the blood or control
solution sample.
The temperature
during the test was
above or below the
operating range.
Move to an area where the
temperature is within the
operating range for blood
glucose test (5–45 ˚C) or the
operating range for blood
β
-ketone test (10–40 ˚C) and
repeat the test after the
meter and test strips have
reached a temperature
within the operating range.
The blood sample
has abnormally
high viscosity or
insufficient volume.
Repeat the test with a new
test strip.