Influence Quantities
Glucose
Packed Cell Volume (Hematocrit)
Packed cell volume evaluation was conducted in various
hematocrit levels. The range of hematocrit levels within the
acceptance criteria is 15–65 %.
Precision:
Precision studies were performed in a laboratory using
the CareSens Dual Blood
β
-Ketone Monitoring Systems.
This study shows that there could be a variation of up to 4.4 %.
Slope
1.012
Y-intercept
-0.0223 mmol/L
Correlation coefficient (r)
0.996
Number of subjects
100
Range tested
0.015–7.2 mmol/L
Within Run Precision
Blood average
0.5 mmol/L
SD = 0.04 mmol/L
1.4 mmol/L
SD = 0.05 mmol/L
3.2 mmol/L
CV = 2.9 %
5.5 mmol/L
CV = 3.1 %
7.2 mmol/L
CV = 3.8 %
Between Run Precision
Control average
0.7 mmol/L
SD = 0.05 mmol/L
2.4 mmol/L
CV = 4.0 %
4.7 mmol/L
CV = 4.4 %