Understanding Error Messages
A used test strip was inserted.
> Repeat the test with a new test strip.
The blood or control solution sample was
applied before the symbol appeared.
> Repeat the test with a new test strip and wait
until the symbol appears before applying
the blood or control solution sample.
The blood sample has abnormally high viscosity
or insufficient volume.
> Repeat the test with a new test strip.
The temperature during the test was above or
below the operating range.
> Move to an area where the temperature is
within the operating range for blood glucose
test (5–45˚C) or the operating range for blood
β
-ketone test (15–30˚C) and repeat the test
after the meter and test strips have reached a
temperature within the operating range.