filtering.
5. If there is a need for contact force monitoring, you can contact the aftersales personnel
to change the IO alarm mode.
6. In the contact force monitoring mode, the resetting will be performed automatically after
the contact force disappears.
7. In the contact force monitoring mode, the contact force threshold is affected by filtering,
and touches by mistake can be avoided.
5.1.6 One-time acquisition mode function
Steps:
1. Connect the host software to the sensor according to 5.1.1.
2. Click the
Stop
button to stop the data acquisition.
3. Check
Single Shot
, and then click
Connect
to enable the one-time acquisition mode of the
sensor.
4.
You can check the frame rate in the lower part of the “Plot” interface to compare the frame rate
differences in different communication modes and filtering.
5.1.7 Crosstalk compensation function
Steps: 1. Connect the host software to the sensor according to 5.1.1.
2. Click the
Stop
button to stop the data acquisition.
3. Check
Crosstalk correct Enable
, and then it will take effect immediately and the crosstalk
compensation mode will be enabled (which is enabled by default for the factory settings).
4.
Click the
Save user setting
button on the
Advanced setting
menu, and then the relevant
parameters will remain valid after shutdown.