Remote Monitoring/Data Logging
The 8075A analog output jack can be used to monitor or log
dew point over a period of time.
The output jack is a 3 pin male M8 type connector. This
connector accepts standard threaded or snap fit female M8
cables.
Pin No.
Wire Color*
Signal
1
Brown
4-20mA (current sourcing)
3
Blue
Ground
4
Black
0-5V
* standard wire color code of M8 cable assemblies
Note that the output scaling is -40
F to +70
F.
1
3
4
Connector pinout
(front view)