40.0
℃
Adj
4.5.3 After 20 min., Press the
and
keys
simultaneously,
the display temperature (in this
example) will show 20.5
℃
, and then begin to rise
to 40.0
℃
automatically
.
During this time, the
*
sign will flicker. When the display temperature
reaches 40.0
℃
, the signs ADJ and
*
will flicker
together. Let the instrument incubate at 40.0
℃
for 20 min.
4.5.4 After 20 min, the thermometer temperature can
be matched with the display temperature. Read
the actual temperature of the thermometer (in this
example) it is 38.8
℃
. Press the
and
keys to
amend the display temperature value to 38.8
℃
,
and then press the
“start” key to confirm.
4.5.5 At this point the incubator temperature will begin
to rise automatically until it reaches 80.0
℃
,
the
*
sign on the display will flicker. When the
display temperature reaches
80.0
℃
, the signs
ADJ and
*
will flicker together. Let instrument
Incubate at 80
℃
, for 20 min.
10
38.8
℃
Adj
80.0
℃
Adj
”
”
Press “Start”