4.5.6 After 20 min, the thermometer temperature
can be matched with the display
temperature. Read the actual temperature
of the thermometer (in this example) it is
79.0
℃
. Press the
or
key to amend the
display value to 79.0, then press the
“start”
key to confirm.
4.5.7 The screen will return to the display as
seen to the right.
After temperature calibration, the
temperature displayed is now the same as
the temperature of the block.
4.5.8.
Canceling Calibration
During temperature calibration, press
and
key simultaneously
to cancel the calibration. If canceled, the system will keep the
former calibration.
11
79.0
℃
Adj
P1 74.5
℃
NOTOK
Press “start”
”
”