Operation Manual 

 

Small Format Digital Dry 

Block Incubator 

Catalog No. INCUBATOR 

 

  

 

  

  

                         

 
 
 
 
 

                                                   

 Rev. C. April 2016

 

 

Summary of Contents for IB003

Page 1: ...Operation Manual Small Format Digital Dry Block Incubator Catalog No INCUBATOR Rev C April 2016...

Page 2: ...UBATOR from Hygiena Opening Check Please check the instrument when you first open the instrument packing case If you find there is something wrong or missing with the instrument contact Hygiena at www...

Page 3: ...ical data 4 3 Safety precautions 5 4 Operation guide 6 4 1 Control keys 6 4 2 Create a new program 6 4 3 Select minute or second for timer 7 4 4 Run and stop a program 7 4 5 Temperaturecalibration 8 5...

Page 4: ...re manual thoroughly before using the incubator 1 1 What is in the box When you unpack the incubator make sure you have the following parts No Item Type QTY 1 Incubator INCUBATOR 1 2 Power adapter and...

Page 5: ...Power switch Power connector 1 2 Incubator diagram 2 Block Display Panel...

Page 6: ...3 2 Connect the incubator to the power adapter connect the adapter to the power cord and connect the power cord to a main power supply 100 240V 1 3 3 To power ON the incubator turn the power switch to...

Page 7: ...y DC12V Power 35 W Temperature range RT 5 80 Timing range 1sec 999 sec or 1min 999 min Accuracy of the temperature 0 5 Display accuracy 0 1 Heating time from 20 to 80 12min Ambient temperature 5 35 Di...

Page 8: ...low temperature little dust no water and away from the sun or heat lamp The area should have good airflow no corrosive gas or a strong disturbing magnetic field It should be far away from central heat...

Page 9: ...temperatures and times for each program The single symbol is the first temperature and time the double symbol is the second temperature and time Press the key to all the way to the right end of the di...

Page 10: ...ram 4 4 1 Press the Prog key to select the program The instrument starts to heat automatically according to the previously set temperature 4 4 2 It shows NOTOK on the display as it is heating up When...

Page 11: ...e temperature of the instrument has been calibrated at the factory If there is deviation between the actual temperature and the displayed temperature the calibration steps in section 4 5 1 may be foll...

Page 12: ...35 In this example an ambient and set program display temperature of 20 5 is used 4 5 2 Inject olefin oil mineral oil into one of the cone shaped wells and then put a thermometer into the well The pr...

Page 13: ...n the thermometer temperature can be matched with the display temperature Read the actual temperature of the thermometer in this example it is 38 8 Press the and keys to amend the display temperature...

Page 14: ...0 then press the start key to confirm 4 5 7 The screen will return tothe display as seen to theright After temperature calibration the temperature displayed is now the same as the temperature of the...

Page 15: ...Instrument Errors Error Cause Solution No display No main power connection Power failure Plug in main cable on both sides Check the fuse OPEN in the display with alarm Broken sensor or loose contact...

Page 16: ...UBATOR For Reference Only Questions Visit www Hygiena com for more information 1 888 HYGIENA Americas 44 0 1923 818821 International 13 100 240V Adapter 12V 4A Mains switch J8 J10 Sensor J3 Main board...