New Q7 Cycler and Computer Package
1. Power on the unit and computer package
a. Cycle startup lights follows – Error, Power, In-Use, Error, Power
i. If error stays on there is problem, try powering down and restarting, if that
doesn’t work contact Diagnostics Support.
b. The drawer will unlock when the “in use” light goes off, open the drawer and
immediately remove the plastic block inside the machine. Keep this on site for the
client in the box with the backup drivers, I like to write “backup BAX® supplies” on it.
2. Launch the software and run the system checks to ensure that the machine is in good
working order.
a. Under the Diagnostics menu, select Cycler, then run the Self Check – this will check
the system communication, lamp power, filter rotation, etc. If it passed you are good
to go.
i. If it fails, record the error (also found in the log file if you forget), then reboot
the system and CPU and try again.
ii. If it is a communication error check; try another USB port. If the error
continues, check that USB port is active on the computer and that the USB
cable is connected and in good condition.
iii. If it is a lamp power error, open the door and make sure that the bulb is
properly seated, it can get loose during shipping. The prongs should remain
level and the bulb should be all the way down.
b. Under the Diagnostics menu, select Cycler, then select Block Control and select the
Raise option and click Run. After that completes, select Lower and Click run. The “in
use” light will go off and the drawer will be able to open.
i. If this fails, reboot the system and try again. If it continues to fail, contact
Diagnostics Support.
c. Under the Diagnostics menu, select Thermal Test, make sure the box for the lid is
check marked and run the test. This will take a few minutes.
i. A passed thermal test will show a stepped pyramid shape with a flat top, the
lines may not be smooth, this is just fine.