Measured values
HMG 4000
Edition 2016-03-08 - V01 R04
[46]
Part No.: 669950
J1939 message
In this mode, the data at a J1939
bus is taken from a transferred
parameter group.
The parameter groups here need to be
transferred directly.
(BAM and CMT cannot be evaluated!)
In J1939 PGN mode, the following
settings are required:
Name
The entered name is used as the name of the measurement channel. If no name is
entered here, "
CAN channel x
" will be shown as the channel name.
Measurement range
Enter the lower and upper measurement range limit, unit and decimal format.
Resolution
Enter the value for the unit / bit.
Offset
Enter any offset there may be.
Priority
Enter the priority of the message (0 - 7).
Datapage
Enter the values for EDP and DP of the PGN (EDP=0 / DP=0 - EDP=1 / DP=1)
PDU format
Enter the value for the PDU format (0 - 255).
PDU specific
Enter the value for PDU Specific (0 - 255).
Source address
Enter the value for the source address (0 - 255).
Bit position
The bit position is entered as a decimal figure with one decimal point. With J1939,
counting starts at 1 (not 0). For this reason only those numbers are allowed which
have figures between 1 and 8 in front of and after the decimal point. (e.g. 2.1 = 2nd
byte / 1st bit)
Bit length
Enter the corresponding bit length of the data (1 - 64).