Safety Information
Proper/Designated Use
Use the unit only for the application described in the following.
The CTU is an extraction unit specially designed to analyze the surface
cleanliness of components by taking wet samples. Analyzing the type, size
and quantity of contamination enables quality standards to be verified and
documented, and the requisite optimization measures to be implemented.
Any use extending beyond or deviating therefrom shall not be considered
intended use. HYDAC FILTER SYSTEMS GMBH will assume no liability for
any damage resulting from such use.
Proper/designated use of the product extends to the following:
•
Maintaining adherence to all the instructions contained herein.
•
Not operating the CTM unless cover panels are installed.
•
Performing requisite inspection and maintenance work.
•
Operation with an impermissible test liquid only.
NOTICE
Test liquids that cannot be used
The CTM will be damaged
►
Only use test liquids which are compatible with the installed materials
and the sealing materials.
Depending on the version (see model code), the CTM may only be used for
the following media (see model code):
CTM-xx 3xx
0
Only use G 60 Spezial as the test liquid
CTM-xx 3xx
1
Only use G 60 Spezial as the test liquid
or
water containing tensides (permissible pH values:
6 … 10) as a test liquid can be used
The use of demineralized water is not permitted.
CTM-EF 3xxx series
en(us)
Page 12 / 48
BeWa CTM-EF 3xxx 4229881a en-us 2018-03-06.docx