Reading the analog output
NAS + T
The NAS+T signal consists of 5 measured values which are transmitted time-
coded with the following time slices:
_
Ti
m
e
Signal
Variable
Signal duration
per pulse
in ms
Current (I) / Voltage (U)
1
Identifier
2 µm
300
High / Low
2
Measured
value
2 µm
3000
Current for measured value
3
Identifier
5 µm
300
High / Low / High / Low
4
Measured
value
5 µm
3000
Current for measured value
5
Identifier
15 µm
300
High / Low / High / Low / High / Low
6
Measured
value
15 µm
3000
Current for measured value
7
Identifier
25 µm
300
High / Low / High / Low / High / Low / High / Low
8
Measured
value
25 µm
3000
Current for measured value
9
Identifier
T
300
High / Low / High / Low / High / Low / High / Low /
High / Low
10
Measured
value
T
3000
Current for measured value
4,0
2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,75
19,5
9,9
10,0
9,85
19,7
0,0
19,8
ContaminationSensor CS1000
en(us)
Page 97 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx
2017-09-25