background image

13

In measuring, if the length of test line is not enough, you can

use lead having same size diameter to extend it.

3. Measurement

Well connected the test line as fig. 2, closed power switch,

power indicator lights, and press the "start/stop" button, then the

display shows "E0000". According to the measured value select

appropriate resistance gear range, press its switch, then its

indicator lights, the four and half bits digital stably shown in LCD is

the measured resistance value. When reading resistance value,

please pay attention to the unit of selected range (mΩ / Ω / kΩ).

Measuring large inductance such as power transformer, you’d

better keep the coils not measured short-circuited to avoid

electromagnetic interference.

4. Discharge

Finished measurement, you can directly disconnect main

power. For measuring inductive object, it is not recommended

immediately disconnect the wiring after disconnected the power,

the instrument should be left to continue discharge. General

discharge time is at least 20 seconds; otherwise the EMF of the

inductance could endanger personal safety.

Summary of Contents for HTZZ-2A

Page 1: ...anual You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column m...

Page 2: ...r relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid...

Page 3: ...uire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the prod...

Page 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Page 5: ...Contents I Overview 6 II Application 7 III Performance characteristics 7 IV Specifications 8 V Panel Layout 10 VI Working Principle 10 VII Operation Methods 11 VIII Problems and Solutions 14 IX Notes...

Page 6: ...In order to change the above situation our company has developed this product This instrument is a high precision regeneration product used to replace DC single double bridge Equipped with advanced sw...

Page 7: ...d cable resistance welded joint resistance of power transformers large motors and instrument transformers III Performance characteristics 1 Fast test The maximum output charge current of the instrumen...

Page 8: ...ccuracy but also makes the instrument have good repeatability 3 Wide measurement range resistance measurement range 1 20K wide range 4 AC and DC power supply Attached rechargeable battery the instrume...

Page 9: ...Resolution 1 5 Constant current source 2A 1 20m 20 m 200m 0 2 2 200mA 2 20 10mA 20 200 1mA 200 2k 100uA 2k 20k 6 Working Voltage DC 12V 16 8V AC 220V 7 Power Dissipation 15W 8 Dimension 365 330 180mm...

Page 10: ...tatus 8 start stop 9 power switch 10 power indicator 11 220V plug 12 ground VI Working Principle There is a constant current source in the instrument that can supply DC current When measuring resistan...

Page 11: ...pply test Closed power switch power indicator will light Press start stop buttons then select appropriate resistance gear switch to do the test Finished testing press start stop button discharge indic...

Page 12: ...tery is charging If rechargeable battery s power is sufficient full indicator will light When using AC power to test the instrument also can charge up the battery Instrument equipped with charging pro...

Page 13: ...n in LCD is the measured resistance value When reading resistance value please pay attention to the unit of selected range m k Measuring large inductance such as power transformer you d better keep th...

Page 14: ...d not press Start Stop key IX Notes 1 After testing please charge the battery in time 2 During DC test if the undervoltage indicator light became bright please use AC power instead of DC power for tes...

Reviews: