--------------------�SLIT LAMP HS-5000 2X 37
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as dete�-mined by an electromagnetic site
s
u
rvey
,
should be less than the compliance level in
b
each frequency range.
lnterference may occur in !he vicinity of
equipment marked with !he following
symbo
l
:
NOTE 1) Al 80MHz and 800MHz, !he higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
oeoole.
Field strengths from f1xed transmitters, such as base stations for radio
(cellularlcordless) telephones and land mobile radios, amateur ra
di
o
,
AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with acc
u
racy
.
To assess the electromagnetic environment dueto fixed RF transmitters, an
electromagnetic srte survey should be co
n
s
id
ered
.
lf the measured field strength in
the location in which !he EUT is used exceeds the applicable RF compliance level
above, the EUT should be observed to verify normal operation. lf abnormal
performance is observed, addltional measures may be necessary, such as re
�rienting or relocating !he EUT.
Over !he frequency range 150kHz to 80MHz, field strengths should be less than
[
V
1
]
Vlm.
37/46
Summary of Contents for HS-5000 2X
Page 21: ...SLIT LAMP HS 5000 2X 21 Figure 2 Composltlon 1 1 21 46...
Page 22: ...22 SLIT LAMP HS 5000 2X U D _ 1 Figure 3 Comoosltlon I1 J 22 46...
Page 26: ...26 SLIT LAMP HS 5000 2X 1 Floure 6 Assemblino the lnstrument II J 26 46...