11.2. Guidance and manufacturer's declaration - electromagnetic
immunity
The EUT is intended far use in the eleclromagnetic environment specified be
l
ow
.
The customer or the user of the EUT should assure that it is used in such an
environment.
lmmunity test
IEC 60601
Compliance
Electromagnetic
Test level
level
environment -ouidance
Electroslatic
±6kV Contact
±6kV Contac t
Floors should be wood,
discharge (ESO)
±BkV air
±BkV air
concrete or ceramic lile.
IEC 61000-4-2
lf floors are covered with
synlhetic material, the
relative humidity
should
be at least 30%.
Electrical fast
±2kV far power
±2 kV far powe r
Mains power
quality
tran si en 1/bu rst
supply lines
supply lines
should
be
thal
of
a
IEC 61000-4-4
±
1kV
far
± 1kV far
typical commercial
or
input/output
input/output
hospital environment.
lines
lines
Surge
±1kV
±1 kV differen tial
Mains
power
quality
IEC 61000-4-5
differential
mode
should
be
that
of
a
mode
±2kV
common
typical
com mercial
or
±2kV
common
mode
hospital environment.
mode
Voltage
dips
,
<5% UT
<5% UT
Mains
power
quality
short
(>95%
dip
in
(>95% dip in UT)
should
be
that
of
a
interruptions and
UT)
far 0.5cycle
typical
commercial
or
voltage
far 0.5cycle
40% UT
hospital environment.
lf
variations
40% UT
(60% dip in UT )
the
user
of
the
EUT
on power supply
(60%
dip
in
far 5 cycle
image intensifier require
input lines
UT)
70% UT
continued
operalion
IEC 61000-4-11
far 5 cycle
(30% dip in UT)
during
power
mai ns
70% UT
far 25 cycle
interruptions,
(30% dip in UT)
<5% UT
it is
recommended that
far 25 cvcle
(<95%
dio
in
the
34/46
Summary of Contents for HS-5000 2X
Page 21: ...SLIT LAMP HS 5000 2X 21 Figure 2 Composltlon 1 1 21 46...
Page 22: ...22 SLIT LAMP HS 5000 2X U D _ 1 Figure 3 Comoosltlon I1 J 22 46...
Page 26: ...26 SLIT LAMP HS 5000 2X 1 Floure 6 Assemblino the lnstrument II J 26 46...
Page 42: ...42 SLIT LAMP HS 5000 2X 12 Electric Diagram 42 46...