![Huvitz HLM-7000P Service Manual Download Page 51](http://html1.mh-extra.com/html/huvitz/hlm-7000p/hlm-7000p_service-manual_2190265051.webp)
50
A
UT
O
-
LE
NSME
T
E
R
4.5.
UV MODULE TEST
Check UV-LED states
UV-LED values in UV mode should change(100% -> 0%).
You can get UV values at most 100%.
(When you get in the UV mode, then popup message shows to adapt auto 100%
function)
Check1
. Check if light radiated from OPTIC
LED comes out from UV measurement part.
NO
Change UV-LED
PCB ASS’Y
Check1
.
Check if UV SENSER in UV measurement part is hidden by
obstacles such as FILTER COVER.
NO
Check1.
Turn on the power switch and enter UV MODE.
Check2
. Turn variable resistance in UV-REC PCB with cross screw driver.
Check3
. Press CAL BUTTON among MODE BUTTON and you can get 100%.
Change UV MODULE
NO
Change MAIN BOARD PCB ASS’Y
Summary of Contents for HLM-7000P
Page 1: ......
Page 2: ...Auto Lensmeter HLM 7000P 1 Service Manual Auto Lensmeter HLM 7000P...
Page 23: ...22 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Page 25: ...24 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Page 27: ...26 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Page 29: ...28 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Page 31: ...30 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Page 33: ...32 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Page 35: ...34 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Page 37: ...36 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Page 39: ...38 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Page 41: ...40 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Page 43: ...42 AUTO LENSMETER 3 5 UV Bone Assembly P I c t u r e...
Page 45: ...44 AUTO LENSMETER 3 6 Front Cover Assembly P I c t u r e...
Page 48: ...Auto Lensmeter HLM 7000P 47...