Auto Lensmeter HLM-7000P
15
2.6.
PD Setup
With the PD Sensor, we can calibrate the PD value of our machine.
The procedures are as followings:
A.
Press the ‘PD’ (
) button.
B.
Push in the PD sensor is the leftmost. And press the ‘L Lim’ button.
C.
Push in the PD sensor is the rightmost. And press the ‘R Lim’ button and check if
‘PD POSITION[0] ~ PD POSITION[8]’ values are changed.
D.
Finally the ‘SAVE’ button to save.
The information displayed in ‘setting the position of PD’ screen is as follows:
Summary of Contents for HLM-7000P
Page 1: ......
Page 2: ...Auto Lensmeter HLM 7000P 1 Service Manual Auto Lensmeter HLM 7000P...
Page 23: ...22 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Page 25: ...24 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Page 27: ...26 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Page 29: ...28 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Page 31: ...30 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Page 33: ...32 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Page 35: ...34 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Page 37: ...36 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Page 39: ...38 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Page 41: ...40 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Page 43: ...42 AUTO LENSMETER 3 5 UV Bone Assembly P I c t u r e...
Page 45: ...44 AUTO LENSMETER 3 6 Front Cover Assembly P I c t u r e...
Page 48: ...Auto Lensmeter HLM 7000P 47...