12
HDC-9100 SERVICE MANUAL
13. The device should neither be kept nor installed in the place with high humidity. For the optimal operation, the
humidity should be at the range of 30%~90%.
14. The manufacturer is not liable for damage caused by unauthorized tampering with the device(s). Such tampering
will forfeit any rights to claim under warranty.
15. To avoid injury by tipping over, set the monitor to the following tilted position before moving.
Below 10°
Below 10°
16. Before connecting the AC power cord to the DC adapter outlet makes sure the voltage designation of the DC adapter
corresponds to the local electrical supply.
17. Equipment grounding is vital for safe operation. Plug power cord into a properly earthed mains supply outlet whose
voltage and frequency characteristics are compatible with those listed on the monitor or in this manual. Do not uses
plug adapters or extension cords; such devices defeat the safety ground and could cause injury.
18.
Grounding reliability is achieved only when the monitor is connected to a “hospital only” or “hospital grade”
receptacle. Inspect routinely and do not use if damage is discovered.
19. Do not touch signal input, signal output or other connectors, and the patient simultaneously.
20. Plug the female and of the AC Power Cord into the AC inlet of the DC Adapter.
Caution: Make sure the power cord is the correct type that is required in your area.
This LCD monitor has a universal power supply that allows operation in either 100-120V AC or 200-240V AC voltage
areas (no user adjustment is required).
Use the proper power cord with correct attachment plug type. If the power source is 120 V AC, use a power cord
which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125 volts AC with UL and C-UL
approvals. If the power source is a 240 V AC supply, use the tandem (T blade) type attachment plug with ground
conductor power cord that meets the respective European country’s safety regulation.
The hospital-grade plug for medical products intended for use in Denmark has DEMKO approval and is rated 13
amps at 250 Vac. Plug is recommended for use in medical applications and specification are being added to the
standard SB 107-2-
D1. Plug mates with maker’s Danish hospital-grade socket. Hospital sockets have slightly