---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 83
(example)
F
1
=
+10D,
d
=
+3mm
:
F
2
=
+9.71D
F
1
=
+10D,
d
=
-3mm
:
F
2
=
+10.30D
F
1
=
-10D,
d
=
+2mm
:
F
2
=
-9.80D
F
1
=
-10D,
d
=
-2mm
:
F
2
=
-10.20D
The following four kinds of data can be chose according to examinee in ‘User Setup Mod
e’.
VD
=
0.0mm: hard contact lens
VD
=
12.0mm: lens of glasses (oriental people)
VD
=
13.5mm: lens of glasses (western people)
VD
=
15.0mm: lens of glasses (western people)
(3) Cylinder Form
There are three kinds of cylinder form.
-
: A form displaying the data using minor meridian. Refraction power of the minor
meridian is displayed as SPH and the axis as AX. The difference between the m
ajor and minor meridian is displayed as –CYL.
+
: A form displaying the data using major meridian. Refraction power of the major
meridian is displayed as SPH and the axis as AX. The difference between the m
F
1
: Refraction Power with the standard VD in diopter
F
2
: Refraction Power with changed VD in diopter
Summary of Contents for CRK-1
Page 1: ...Service Manual Auto Ref Keratometer CRK 1...
Page 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Page 35: ...34 Auto Ref Keratometer CRK 1...
Page 41: ...40 Auto Ref Keratometer CRK 1...
Page 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Page 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Page 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Page 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Page 76: ...Auto Ref Keratometer CRK 1 75...
Page 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...