---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 55
D. If the cross stopped at the center of mire ring without passing, then the test is
success.
3) If Step 2. is failed, increase setting value and retry.
4) If Step 2. is succeed current values are the answer.
5) Repeat these process to the other side.