---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 53
4.2. Auto Tracking Setup
If auto tracking power is too big or too weak, you can adjust coefficients to control the tr
acking power.
Menu for tracking coefficients control is “2-2. TRACKING SETUP”
Bottom two items are for auto tracking.
Others are for joystick scrolling, and they are not for you. You do not need to touch thos
e items for any cases.
We recommend following procedure to find appropriate coefficients.