![Huvitz CLM-4000 Service Manual Download Page 66](http://html1.mh-extra.com/html/huvitz/clm-4000/clm-4000_service-manual_2190259066.webp)
Auto Lensmeter CLM-4000
65
4.6.
PD TEST
z
PD value is not displayed
⋯
z
PD value is so strange
⋯
Check1. Check right/left switching when you move LEVER
right and left compared to the center of PD.
NO
Change MAIN
Change PD MODULE
Y
E
S
Check1
.
Check that PD LEVER is in initial position before you turn the
power switch.
Y
E
S
Check1. First, turn off the POWER switch and place PD LEVER in initial
position. And then turn on the power switch.
Information about initial position of PD
LEVER: After moving PD LEVER as left
as you can (initial position), turn on the
power switch.
Summary of Contents for CLM-4000
Page 1: ......
Page 2: ...Auto Lensmeter CLM 4000 1 Service Manual Auto Lensmeter CLM 4000...
Page 5: ...4 AUTO LENSMETER 1 Introduction 1 1 Components List Figure 1 Components Names I...
Page 6: ...Auto Lensmeter CLM 4000 5 Figure 2 Components Names II Figure 3 Components Names III...
Page 26: ...Auto Lensmeter CLM 4000 25 Center Point A12 Angle of two points...
Page 29: ...28 AUTO LENSMETER 3 Repair Standard 3 1 Removing side cover assembly P I c t u r e 1 2 3 4 5...
Page 31: ...30 AUTO LENSMETER 3 2 Removing front cover Upper ass y P I c t u r e 1 2 3 4 5 6 7 8...
Page 33: ...32 AUTO LENSMETER 3 3 Removing Printer ass y P I c t u r e 1 2 3 4 5 6...
Page 35: ...34 AUTO LENSMETER 3 4 Removing PD assembly P I c t u r e 1 2 3 4 5 6 7 8 9 10...
Page 39: ...38 AUTO LENSMETER 3 5 Removing Body front cover Lower ass y P I c t u r e 1 2 3...
Page 41: ...40 AUTO LENSMETER 3 6 Removing UV assembly P I c t u r e 1 2 3 4 5...
Page 43: ...42 AUTO LENSMETER 3 7 Removing Movement assembly P I c t u r e 1 2...
Page 45: ...44 AUTO LENSMETER 3 7 1 Removing Movement Left Subassembly P I c t u r e...
Page 47: ...46 AUTO LENSMETER 3 7 2 Removing Movement Right Subassembly P I c t u r e...
Page 49: ...48 AUTO LENSMETER 3 8 Removing Gear assembly P I c t u r e 1 2 3 4 5 6 7 8...
Page 51: ...50 AUTO LENSMETER 3 9 Removing Main board assembly P I c t u r e 1 2...
Page 53: ...52 AUTO LENSMETER 3 10 Removing Back bone assembly P I c t u r e 1 2 3...
Page 57: ...56 AUTO LENSMETER 3 10 2 Removing CCD Camera assembly P I c t u r e 1 2 3 4 5 6 7...
Page 59: ...58 AUTO LENSMETER 3 11 Removing SMPS Assembly and Power cover P I c t u r e 1 2 3 4 5 6...