EN
© 2022 HUSLOG
All rights reserved
Touchless thermometer
16
Guidelines and manufacturer's declaration - electromagnetic resistance
Emitted radio frequency signal IEC 61000-4-3
(Test specification for the CABINET PORT RESISTANCE to RF signal of wireless
communication equipment)
Text frequency
(MHz)
Band
(MHz)
Service
Modulation
Max.
power
(W)
Distance
(m)
Level of
durability test
(V/m)
385
380-
390 TETRA 400
Impulse
modulation
18 Hz
1.8
0.3
27
450
430-
470
GMRS 460,
FRS 460
FM
Band deviation
+/- 5 kHz
Sinusoid wave
1 kHz
2
0.3
28
710
704-787 LTE bands
13, 17
Impulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800-
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Impulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700-
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
bands 1, 3,
4, 25, UMTS
Impulse
modulation
217 Hz
2
0.3
28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN
802.11
b/g/n, RFID
2450, band
LTE 7
Impulse
modulation
217 Hz
2
0.3
28
5240
5100-
5800
WLAN
802.11 a/n
Impulse
modulation
217 Hz
0.2
0.3
9
5500
5785