9. Maintenance
Huntleigh Healthcare recommend the Huntleigh range of patient lifting
devices and lifting accessories receive a thorough six-monthly
examination and test. We recommend the examination and test is carried
out by a competent person. This product should also receive a full service
once a year.
Huntleigh Healthcare can offer this after sales care, please ask for our
Flexi-Serve brochure, which contains details of the full range of service
and maintenance contacts on this product.
This is in line with the requirements of No2307 Health and Safety: The
Lifting Operations and Lifting Equipment Regulations 1998 (LOLER)
The examination, tests, servicing and repairs should be carried out
according to the recommendations detailed in the hoist Service Manual
which is available upon request.
IMPORTANT
During the 12-month warranty period, the electrical
components must not be opened or tampered with, as this
will invalidate the warranty.
WARNING
Use only Huntleigh Healthcare approved parts for repair or
replacements.
First Line Maintenance
(For more detail please refer to Service Manual)
Spreader Bar
Check for:
•
Freedom of rotation and swing.
•
Wear on the spreader bar/jib suspension point
•
Wear on the central pivot.
•
Secure attachment to the jib.
•
Effective function of the safety latch
•
Excessive wear on the sling hooks.
•
Sling strap retainers effective function.
Maintenance – lubricate main suspension point and central pivot. Replace
missing or damaged parts.