41 Replacing the Snap-lock Connectors
Note : Older versions have a pump connectors moulded within the side panel and are not
replaceable.
41.1 Remove pressure control module with side panel (Refer to Section 29.1 to 29.7)
41.2 Refer to Fig 15, disconnect the three signal connectors (Fig 15, Item 30 ) and the power wires
from the PCB.
41.3 Remove screw (Fig 15, Item 10) and remove the PCB.
41.4 Undo locknut and withdraw connector through side panel with tube.
41.5 Remove tube from connector.
41.6 Replace with the same type. Replacement is the reverse of the above procedure. Ensure that
the snap lock release clip on the female (lower) connector is pointing toward the rear of the
case. Make sure the locknut is on the pressure module before the connector and tube is
inserted.
Issue 2
Chap 3
Nov 2001
SER0004
Page 45
Mini Pump Service Manual
Pump Repair
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...