32 Replacing the Neon Indicator Light
Note : The neon indicator light is only applicable to the mechanical versions of the pump. The
electronic version of the pump has two red indicators, a green indicator and a yellow indicator.
These are wired directly into the PCB and are replaced only by exchanging the complete PCB.
32.1 Disconnect the indicator lead from the terminal block (terminal 2) (Fig 14, Item 10).
32.2 Disconnect the indicator lead from the Micro switch (Fig 18, Item 10).
32.3 Release the screw (Fig 17, Item 20) attaching the neon mounting plate (Fig 17, Item 10) and
remove the neon (Fig 17, Item 30) and mounting plate assembly.
32.4 Note the orientation of the components, unclip the neon (Fig 17, Item 30) from the lens and
push the lens out of the mounting plate.
32.5 Replacement is the reverse procedure of the above. Ensure the correct orientation of
components. Connect the wires to the terminal block and the Micro switch in accordance with
the wiring diagram for the mechanical version of the pump unit (Chap. 4, Fig 25).
Issue 2
Chap 3
Nov 2001
SER0004
Page 33
Mini Pump Service Manual
Pump Repair
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...