The power failure alarm is initiated by the removal of the mains/power supply from the unit for a
period of more than 5 seconds, by a means other than operation of the ‘Run/Standby’ switch. A drop
in the supply voltage below the minimum supply specification will be interpreted as a supply failure. If
power is restored within 5 seconds, no alarm is given and normal operation is resumed.
A battery backup supply provides power to operate the visual and audible alarm.
Battery Management (Electronic Units Only)
On switching the pump to ‘Run’, the processor automatically initiates a standard charge sequence
(3 mA) for a predetermined period. On completion of the standard charge period, a trickle charge
(0.9 mA) will be applied to the battery for the remainder of the operating period.
When discharging during a power failure alarm, the battery supply will be shut down in a controlled
manner if the terminal voltage falls below 4.6 V before the alarm duration has expired.
Issue 2
Chap 1
Nov 2001
SER0004
Page 7
Mini Pump Service Manual
Introduction
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...