Aeros Supplemental Manual for EasyMatch QC ver. 2.0
20
Figure 19. Standardize
Attach the standardization box to the sensor. Then attach the black glass to the standardization box. Press
NEXT
to continue.
Figure 20. Read the Black Glass
Remove the black glass and attach the white tile. Press
NEXT
to continue.
Figure 21. Read White Tile
Attach the green tile and press
NEXT
to continue. Ten readings of the green tile are made and
averaged. The average reading and its Pass/Fail status will be shown. Confirm that the result is
PASS
.
The Pass/Fail tolerance for the Aeros is ± 0.30.
Summary of Contents for Aeros EasyMatch QC
Page 2: ...Aeros Supplemental Manual for EasyMatch QC ver 2 0 2 ...
Page 9: ...Aeros Supplemental Manual for EasyMatch QC ver 2 0 9 ...
Page 17: ...Aeros Supplemental Manual for EasyMatch QC ver 2 0 17 ...
Page 28: ...Aeros Supplemental Manual for EasyMatch QC ver 2 0 28 ...
Page 30: ...Aeros Supplemental Manual for EasyMatch QC ver 2 0 30 ...