15
Pump Modes of Operation:
The main water pump has three options for controlling the velocity of wa-
ter flowing over the sample. Depending on the water flow speed desired;
automatic, manual, or sump control may be chosen for managing the
water velocity for the test.
Automatic Mode:
In Automatic Mode manages the water flow speed between the ranges
of 0.15 m/s to 7.0 m/s. When in Automatic Mode, the controller will auto-
matically monitor the water velocity and adjust the pump’s water output
to maintain the target water velocity. This the default control setting for
the EFA, and nothing need be done in order to operate the EFA in Auto-
matic Mode.
Flow Controls:
1
1. To use the Automatic Mode simply enter the desired water velocity on
the touch screen and press the FLOW button.
Summary of Contents for HM-5940
Page 1: ...product manual 4 22 HM 5940 Erosion Function Apparatus ...
Page 5: ...5 Figure 2 Elevation View of EFA Figure 3 Plan View of EFA ...
Page 6: ...6 Figure 4 Detail View of Sample Compartment and Function of Components Within ...
Page 25: ...25 This screen is accessed by clicking on the Preferences button 1 1 ...