Notes/Caution

• 

Do not attempt to tighten screws on or disassemble the FD-4A unit, replace parts, 
adjust the transformer core, make internal adjustments to the DTR-451 or perform 
calibration routines other than those outlined above. Doing so may necessitate 
factory recalibration and/or repair

• 

The FD-4A unit, LVDT and DTR-451 digital readout unit function inter- dependently. 
All are serially numbered, individually calibrated and must remain together as a 
matched set Do not shorten or lengthen the cable from the FD-4A to the DTR-451. 
Components cannot be interchanged from one unit to another or ordered separately.

Note:

 Because each DTR-451 unit is calibrated to a specific FD-4A adapter do not 

contact Humboldt Mfg. Co. directly if a problem develops with the readout unit. Repairs 
must be directed through the factory so components can be recalibrated after the repair.

• 

Locate the DTR-451 unit in a protected area so water from the ductilometer bath 
splashed onto it. Allow space around the unit for heat dissipation. Leave the 
continually. A power conditioning unit is recommended especially if the apparatus is 
an area where nearby industrial machinery creates spikes or surges in the power line. 
Power fluctuations can affect the output to your recording devices.

• 

Do not immerse the FD-4A beyond the marking on the base. A one inch safety 
margin is allowed to protect the LVDT. Although the LVDT coil is vacuum 
impregnated with varnish and has a stainless steel casing, it is not a submersible 
design. Permanent damage may result if brine comes in contact with it for any period 
of time.

Note:

 While the FD-4A adapter is immersed in a cold bath, water from the air may 

condense on the upper portion of the body where the LVDT is located. High humidity 
and/or condensation will not adversely affect the transformer or core.

• 

Securing the calibration stand to a workbench or table is required. The lightweight 
phenolic base will tip forward in the higher weight ranges. Always use the cable and 
weight holder supplied with the unit to hold your weights. Do not exceed 30 pounds 
of weight for calibrating purposes.

• 

Do not clean the calibration stand with strong solvents such as methyl ethyl ketone. 
Do not immerse the stand in solvent or water.

• 

Be careful when removing weights from the weight holder. The cable will untwist 
somewhat as weights are applied and it must be allowed to twist back as weights are 
removed. If all 30 pounds are removed suddenly, the cable may coil and/or kink.

Data Acquisition

To gain full advantage of the FD-4A apparatus, some type of data acquisition system is 
required. Although tests can be run with a strip chant recorder or other output device, 
computerized data acquisition has the advantage of being able to analyze, calculate, 
compare, store and present data gathered from tests according to user requirements.  
It is fast, accurate, and flexible.  
Data acquisition cards and matching software packages which are compatible with 
various personal computers and existing spreadsheet programs are available from 
numerous sources. Many will work with the analog output from the DTR-451.

CAUTION:

 Keep hands, clothing and other objects away from moving parts when the 

machine is in operation.

Summary of Contents for FD-4A

Page 1: ...product manual 07 09 H 1021 FORCE DETERMINATION ADAPTER AND DIGITAL INDICATOR ...

Page 2: ......

Page 3: ...ctronic components are safely located out of the water bath The digital transducer readout unit incorporates an analog output for connection to a strip chart recorder or data acquisition unit Once started tests can be monitored unattended so technicians are free for other work The stainless steel construction eliminates rust and corrosion Each FD 4 unit is serially numbered and individually calibr...

Page 4: ...is a self contained power line operated high stability LVDT signal conditioner with a 4 digit readout It supplies a high frequency AC voltage for the LVDT and a 0 to 3 0 VDC analog output for data acquisition interfacing A switch on the front panel can change the display from pre scaled English units to pre scaled metric units Calibration Stand This one piece phenolic block has a ledge and stainle...

Page 5: ...gital converter computer or other readout device may be used One tenth of one volt output is equal to one pound of force on the adapter Additional information regarding the 15 pin connector and the optional BCD interface if required is available Calibration Set up The FD 4A apparatus is factory calibrated before shipment and is ready to use It is recommended however that the zero adjustment be che...

Page 6: ...ish Metric adjustment do not change analog output voltage Likewise the select switch changes only the display and does not alter the analog output voltage proportionally for the metric scale Add and subtract weights and make adjustments until the readings are within tolerance The output device or data acquisition equipment should also be checked to verify that the analog output voltage corresponds...

Page 7: ...If necessary use the zero adjustment to set the readout and analog output to null Your weight calibrations will remain intact It will be necessary to determine a new starting position for the carriage of the ductilometer This position can be easily found by using the sample and moving the carriage until the specimen slides freely over both the FD 4A platform pin and the pin on the ductilometer Not...

Page 8: ...Permanent damage may result if brine comes in contact with it for any period of time Note While the FD 4A adapter is immersed in a cold bath water from the air may condense on the upper portion of the body where the LVDT is located High humidity and or condensation will not adversely affect the transformer or core Securing the calibration stand to a workbench or table is required The lightweight p...

Page 9: ...steel types 302 303 304 17 7 PH Max submersion depth 2 125 Max tensile stress measurement Cable to DTR unit 30 lbs Length 72 Spring plates Thickness I3GA 0 091 Material 17 7 PH Heat treat cond TH 1050 Transducer and Digital Readout Readout Model DTR 451 LVDT Model 100 HR Height 2 1 Dia 0 812 Width 6 125 Length 1 81 Length 10 0 Weight 2 75 lbs Input from FD 4A 9 pin D subminiature Output 15 pin IC ...

Page 10: ... than routine cleaning of parts with appropriate cleaner Careful handling is recommended to ensure best performance Do not attempt disassembly or repairs No special maintenance is required for the DTR 451 digital readout unit and 100 HR LVDT Maintenance on the calibration stand is limited to cleaning off foreign material that may accumulate and checking the pulley bearing for free rotation The bea...

Page 11: ......

Page 12: ...r parts of such product for the warranty of this product please refer to Humboldt Mfg Co catalog on Terms and Conditions of Sale The purchaser is responsible for the transportation charges Humboldt Mfg Co shall not be responsible under this warranty if the goods have been improperly maintained installed operated or the goods have been altered or modified so as to adversely affect the operation use...

Reviews: