SAMPLE CALIBRATION VALUES

Tolerance = ±1% + 2 counts

POUNDS

KG

TOL LOW

TOL HIGH

1.00

0.454

0.97

1.03

5.00

2.268

4.93

5.07

10.00

4.536

9.88

10.12

15.00

6.804

14.83

15.17

20.00

9.072

19.78

20.22

25.00

11.340

24.73

25.27

30.00

13.608

29.68

30.32

NOTES:

1.  For values not shown above, use tolerance formula at top page.
2.  Display should read between high and low tolerance as closely as possible to the 

value of the calibration weight.

3.  Analog output voltage is 1/10 of the weight. Output voltage must also be between 

high and low values as calculated by tolerance limits for any given calibration 
weight.

Operation

Place the FD-4A adapter on the pin of the stationary ductilometer stand in place of the 
sample briquet. Use caution so as not to immerse the adapter beyond the marking. 
Since the stainless steel spring plates are sensitive to temperature changes, a shift on the 
digital readout maybe noticed when the adapter is immersed, especially if there is a large 
difference between the water bath temperature and the ambient temperature. Allow 
the adapter to stabilize for 30 to 60 minutes before conducting any testing. If necessary, 
use the “zero” adjustment to set the readout (and analog output) to null. Your weight 
calibrations will remain intact.
It will be necessary to determine a new starting position for the carriage of the 
ductilometer. This position can be easily found by using the sample and moving the 
carriage until the specimen slides freely over both the FD-4A platform pin and the 
pin on the ductilometer. Note this position on the indicator and return the carriage to 
this position at the beginning of each test. Now, with the sample in place, testing may 
begin. Occasionally, during course of test, the force exerted on the FD-4A adapter may 
exceed the 30 pound limit. Although this is not recommended, there is at least 10 pound 
overload safety margin where the spring plates will not experience any deformation. 
Before running the next test, be sure to check the zero setting and adjust if necessary. 
If the display cannot be adjusted to zero or the analog output does not correspond 
correctly with the display during calibration, it may be necessary to return the apparatus 
to the factory for recalibration.

Remember:

 If the unit is removed from the ductilometer and set on the calibration 

stand, it must be allowed to acclimate to room temperature before any weight checks 
can be made. Then readjust the zero before beginning.

Summary of Contents for FD-4A

Page 1: ...product manual 07 09 H 1021 FORCE DETERMINATION ADAPTER AND DIGITAL INDICATOR ...

Page 2: ......

Page 3: ...ctronic components are safely located out of the water bath The digital transducer readout unit incorporates an analog output for connection to a strip chart recorder or data acquisition unit Once started tests can be monitored unattended so technicians are free for other work The stainless steel construction eliminates rust and corrosion Each FD 4 unit is serially numbered and individually calibr...

Page 4: ...is a self contained power line operated high stability LVDT signal conditioner with a 4 digit readout It supplies a high frequency AC voltage for the LVDT and a 0 to 3 0 VDC analog output for data acquisition interfacing A switch on the front panel can change the display from pre scaled English units to pre scaled metric units Calibration Stand This one piece phenolic block has a ledge and stainle...

Page 5: ...gital converter computer or other readout device may be used One tenth of one volt output is equal to one pound of force on the adapter Additional information regarding the 15 pin connector and the optional BCD interface if required is available Calibration Set up The FD 4A apparatus is factory calibrated before shipment and is ready to use It is recommended however that the zero adjustment be che...

Page 6: ...ish Metric adjustment do not change analog output voltage Likewise the select switch changes only the display and does not alter the analog output voltage proportionally for the metric scale Add and subtract weights and make adjustments until the readings are within tolerance The output device or data acquisition equipment should also be checked to verify that the analog output voltage corresponds...

Page 7: ...If necessary use the zero adjustment to set the readout and analog output to null Your weight calibrations will remain intact It will be necessary to determine a new starting position for the carriage of the ductilometer This position can be easily found by using the sample and moving the carriage until the specimen slides freely over both the FD 4A platform pin and the pin on the ductilometer Not...

Page 8: ...Permanent damage may result if brine comes in contact with it for any period of time Note While the FD 4A adapter is immersed in a cold bath water from the air may condense on the upper portion of the body where the LVDT is located High humidity and or condensation will not adversely affect the transformer or core Securing the calibration stand to a workbench or table is required The lightweight p...

Page 9: ...steel types 302 303 304 17 7 PH Max submersion depth 2 125 Max tensile stress measurement Cable to DTR unit 30 lbs Length 72 Spring plates Thickness I3GA 0 091 Material 17 7 PH Heat treat cond TH 1050 Transducer and Digital Readout Readout Model DTR 451 LVDT Model 100 HR Height 2 1 Dia 0 812 Width 6 125 Length 1 81 Length 10 0 Weight 2 75 lbs Input from FD 4A 9 pin D subminiature Output 15 pin IC ...

Page 10: ... than routine cleaning of parts with appropriate cleaner Careful handling is recommended to ensure best performance Do not attempt disassembly or repairs No special maintenance is required for the DTR 451 digital readout unit and 100 HR LVDT Maintenance on the calibration stand is limited to cleaning off foreign material that may accumulate and checking the pulley bearing for free rotation The bea...

Page 11: ......

Page 12: ...r parts of such product for the warranty of this product please refer to Humboldt Mfg Co catalog on Terms and Conditions of Sale The purchaser is responsible for the transportation charges Humboldt Mfg Co shall not be responsible under this warranty if the goods have been improperly maintained installed operated or the goods have been altered or modified so as to adversely affect the operation use...

Reviews: