background image

8/10

Special attention to the following is necessary:

 

Installation of the unit: Proper ventilation, levelling of the centrifuge, rigidity and stability of the bench.

 

Rotor installation: Check that the rotor nut is tightened firmly.

 

Cleaning of the accessories and the rotor chamber.

 

Load balancing

 

Samples: The cleaning of the accessories is particularly important when using potentially infectious

materials.

 

&OHDQLQJ

Ideally, the rotor should be washed after every use but at least weekly in warm water containing a few drops of

mild liquid soap (domestic washing liquid is ideal ) and any time after spillage has occurred.

Each rotor insert must be washed thoroughly using a small nylon brush. Do not use metal wire brushes.

Dry the rotor with a soft, absorbent, non- woven cloth or tissue. Drying may be finished off with a warm air jet

(e.g. a hair-dryer).

Make sure that no deposits remain in the bottom of the insert holes because the pressure of a flask or tube from

above during centrifugation will significantly increase the chances of breakage.

 

&RQWDPD]DUGV

Our centrifuges are used in medical research, where hazardous substances, including radioactive chemicals, are

frequently found.

Always use the appropriate decontamination procedures where the rotor is exposed to these chemicals.

Examples of commonly used techniques are outlined below. The information is given as a guide only. It is the

responsibility of the owner to use the most suitable procedure.

The rotor should always be completely disassembled before being subjected to heat and after external chemical

cleaning.

 

'LVLQIHFWLRQ

Alcohol ( 70% ethanol or isopropanol ) applied for 10 minutes is ideal for destroying bacteria and viruses.

Alternatively, rotors can be autoclaved for 20 minutes at 120°C to destroy micro-organisms.

Any part which has been subjected to temperatures above 130°C must be discarded.

 

(OHFWULFDO

High voltage is present behind the panels of the centrifuge.

Caution: Do not open the instrument! There are no user-servicable parts inside.

In the case of a malfunction, please contact your local HUMAN representative.

 

,PSURSHU8VH

 

Use only rotors and accessories designed for use in the centrifuges.

 

Do not attempt to override the lid interlock assembly.

 

Load and lock the rotor only in the recommended way. As the centrifuge starts to spin, an improperly

loaded rotor could cause sufficient force to damage the drive shaft and the rotor chamber.

Summary of Contents for HUMAX 4K

Page 1: ...Gtl w 3J User Manual Cat No 15210 100...

Page 2: ...i Qduhrhnm Khrs ne sgd L mt k No DATE Rev REVISION DESCRIPTION 1 01 2007 01 First edition...

Page 3: ...ii...

Page 4: ...llation of this product This warranty does not apply to damages incurred in shipment of goods Any damage so incurred shall be reported to the freight carrier for settlement or claim 3 INTENDED USE OF...

Page 5: ...at least potentially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the labo...

Page 6: ...sibly contaminated parts Before the instrument is removed from the laboratory for disposal or servicing it must be decontaminated Decontamination should be performed by authorised well trained personn...

Page 7: ...b...

Page 8: ...lock System 4 5 Instructions for Use 5 5 1 Schematic Drawing 5 5 2 Controls and Indicators 6 5 3 Preparing The Run 6 5 4 Programming 7 5 5 Program Recall 7 5 6 Starting Running and End Of Run 7 5 7 Er...

Page 9: ...does not allow the centrifuge to operate or the rotor to spin if the lid is not closed The lid locking system allows the lid to be opened when the program ends by pressing a single key There are audib...

Page 10: ...balance The central hole of the rotor and the motor shaft must also be kept clean and dry at all times RDGLQJ Each tube insert must be the same weight as the one diametrically opposite for balancing C...

Page 11: ...d set step 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 Hz Power Consumption 350 W External Dimensions WxDxH mm 274 x 355...

Page 12: ...5 10 XPDQ X0D User Manual 5 Instructions for Use 6FKHPDWLF UDZLQJ Figure 1...

Page 13: ...o set the speed and time during programming and to check the RCF value during the operation 5 Pulse Key Causes the motor to increase to the maximum speed The motor will continue to spin at that speed...

Page 14: ...otor stops spinning When the program is completed END is shown on the speed display 1 and the warning beep sounds to indicate to the user that program has ended The warning beep will continue to sound...

Page 15: ...ere hazardous substances including radioactive chemicals are frequently found Always use the appropriate decontamination procedures where the rotor is exposed to these chemicals Examples of commonly u...

Page 16: ...DLO No daily cleaning is required except in the case of accidental glass breakage or a large amount of spillage in the bowl HHNO Clean the bowl and the accessories with a cotton wool pad dipped in 70...

Page 17: ...10 10 8 Electrical Circuit Diagram...

Page 18: ...HUMAN Gesellschaft f r Biochemica und Diagnostica mbH Max Planck Ring 21 65205 Wiesbaden Germany Tel 49 61 22 99 88 0 Fax 49 61 22 99 88 100 e Mail human human de www human de 01 2007 01...

Reviews: